New Harmonised Standards Under MDR and IVDR
On 10 November 2026, the European Commission published Commission Implementing Decision (EU) 2026/2479 (MDR) and Commission Implementing Decision (EU) 2026/2481 (IVDR), both dated 9 November 2026, introducing further updates to the lists of harmonised standards supporting the MDR and IVDR.
Both decisions replace EN ISO 11137-1:2015 (radiation sterilisation) with EN ISO 11137-1:2026, while the MDR decision also replaces EN ISO 14155 and adds seven new references.
Key Updates
- MDR – Superseding and Newly Harmonised Standards – Implementing Decision (EU) 2026/2479 amends Decision (EU) 2021/1182, including:
- Superseding standards (replacing previous versions):
- EN ISO 11137-1:2026 (radiation sterilisation of health care products), replacing EN ISO 11137-1:2015 and EN ISO 11137-1:2015/A2:2019
- EN ISO 14155:2026 (clinical investigation of medical devices for human subjects – good clinical practice), replacing EN ISO 14155:2020 and EN ISO 14155:2020/A11:2024
- Newly harmonised standards:
- EN ISO 4074:2026 (natural rubber latex male condoms)
- EN ISO 10993-6:2026 (biological evaluation – tests for local effects after implantation)
- EN ISO 10993-7:2026 (biological evaluation – ethylene oxide sterilisation residuals)
- EN ISO 15883-6:2026 (washer-disinfectors employing thermal disinfection for non-critical medical devices and health care equipment)
- EN IEC 60601-2-22:2020 and EN IEC 60601-2-22:2020/A11:2026 (surgical, cosmetic, therapeutic and diagnostic laser equipment)
- EN IEC 60601-2-57:2026 and EN IEC 60601-2-57:2026/A11:2026 (non-laser light source equipment)
- EN ISO 80369-1:2026 (small-bore connectors for liquids and gases – general requirements)
- Superseding standards (replacing previous versions):
- IVDR – Superseding Standard – Implementing Decision (EU) 2026/2481 amends Decision (EU) 2021/1195:
- Sterilisation:
- EN ISO 11137-1:2026 (radiation sterilisation of health care products), replacing EN ISO 11137-1:2015 and EN ISO 11137-1:2015/A2:2019
- Sterilisation:
Both decisions entered into force on the day of their publication. The references to the previous versions of EN ISO 11137-1 (MDR and IVDR) and EN ISO 14155 (MDR) are deleted from the lists with effect from 10 May 2028, giving manufacturers time to adapt. By applying these standards, manufacturers can benefit from the presumption of conformity with the relevant requirements of the MDR and IVDR (Article 8(1) of both Regulations), making them a key tool for demonstrating compliance.
As the previous versions remain referenced until 10 May 2028, manufacturers should assess their transition timelines and update technical documentation accordingly.
What Does This Mean for You?
Manufacturers should review whether the new or updated harmonised standards apply to their devices, particularly where radiation sterilisation, clinical investigations or the other listed topics are involved. Technical documentation and ongoing projects should be reviewed and updated where necessary to align with the latest harmonised standards, taking the transition until 10 May 2028 into account.