Post-Market Surveillance (PMS) and Post-Market Clinical Follow-Up (PMCF) under the MDR: Ensuring safety and performance

Author Avatar
Petra De Geest & Pia Gyselen
Clinical
Medical Devices

What is post-market surveillance and post-market clinical follow-up for medical devices under the MDR in the EU, and how can QbD help? 

PMS and PMCF Activities for Medical Devices Under the MDR | QbD Group
3:52

With the introduction of the Medical Device Regulation (MDR), the requirements for placing medical devices on the market have significantly increased. Additional manufacturers’ obligations concerning post-market surveillance (PMS), including post-market clinical follow-up (PMCF), have gained importance.  

Under the MDR, PMCF expectations include specific requirements, such as the need for a PMCF plan, specifying the methods and procedures for clinical data collection, and clearly stated objectives for PMCF activities for devices of all risk classes. 

In this blog post, we will explore what PMS and, more specifically, PMCF entail and why they are vital for both manufacturers and users of medical devices.  

Understanding PMS and PMCF under the MDR 

PMS is a comprehensive process that manufacturers must establish to collect and review experiences gained from the medical devices they have placed on the market. The primary aim is to ensure that any potential safety issue is identified and assessed promptly, guaranteeing the continuous improvement of the device’s safety and performance. 

PMCF is a component of the overall PMS. PMCF activities involve systematically and continuously collecting and analyzing real-world clinical data to confirm the safety and performance of the device throughout its entire lifetime. This ensures the acceptability of known risks and helps detect emerging risks or unforeseen adverse events. 

Through PMS and PMCF, manufacturers can: 

  • Monitor the real-world use and performance of their device 
  • Identify and investigate any potential safety issues or adverse events 
  • Gather data to support new indications or improvement of existing products 
  • Fulfill regulatory requirements and maintain compliance with the MDR. 

Various types of PMCF activities 

Several types of PMCF activities can be undertaken by the manufacturer to gather post-market clinical data. Each type has its strengths and limitations, and the choice should be based on a thorough gap analysis of the existing clinical data.

This analysis should determine if the data cover the device’s lifetime, are sufficient in quantity and quality, support each claimed clinical benefit, and comply with current quality and ethical standards.

If any gaps are identified, PMCF activities should be planned. Usually, a combination of multiple PMCF activities is recommended to gather sufficient clinical data to support the MDR requirements. 

 

Published literature 

An ongoing review of the scientific literature related to the medical device or similar/equivalent devices can provide additional evidence of safety and performance, as well as identify new risks or adverse events.
 
This literature review is considered to be the minimum required PMCF activity mandatory for all medical devices of all risk classes and should ideally be documented every year. 

Registry studies 

Patient registries offer an organized, systematic collecting of data on individuals with a specific disease, condition, or exposure. These registries help researchers: 

  • understand disease natural history,  
  • identify risk factors,  
  • evaluate treatment effectiveness, and  
  • monitor safety 
Patient registries can be organized by various entities, including researchers, patient organizations, professional societies, government agencies, or healthcare providers. Medical Device manufacturers may support registries to gather information following the approval of their therapies. 
 
Data from patients using a specific medical device can be systematically collected through these types of registries and subsequently analyzed to assess its performance in a real-world setting, hence providing valuable information on its use as intended without the restrictions of controlled clinical investigations.  
 
However, there are challenges with this type of data registration: the quality of the data can be unpredictable, particularly when dealing with missing data and potential biases. These issues depend on who is entering the data and the types of patients registered in these databases. 

PMCF studies 

Post-Market Clinical Follow-Up studies are systematic investigations conducted with a CE-marked medical device. They aim to collect additional clinical data to assess the device’s safety, performance, and benefits during real-world use. PMCF studies add to pre-market data and contribute to risk management.
 
These studies involve well-designed clinical investigation plans, research questions, and statistical analysis plans. PMCF studies can be prospective or retrospective clinical investigations, but ethical standards are similar to pre-market clinical investigations, ensuring subject protection and requiring an informed consent. 
 
Although this type of post-market clinical data is of the highest value for the medical device in scope, it also comes with several challenges: 
  • A robust data collection system needs to be established to collect accurate and comprehensive data from real-world use.
  • Proper informed consent while respecting patient privacy is essential.
  • Data gathering while minimizing the burden on healthcare providers requires careful planning and coordination.
  • Ethical approval can be time-consuming, and.
  • The cost of this type of investigation can be high. 

Observational clinical investigations 

In Prospective Observational Clinical Investigations, the investigator assigns the subject to an exposure or treatment with a CE -marked medical device, requiring a signed informed consent, and then collects the data from that point forward. Normal clinical practice is followed without extra burden to the patient. 
 
Retrospective Observational clinical investigations (Cohort Reviews) are studies where data are already available throughout the course of the normal clinical practice and documented in the medical records of the patient. The investigator collects the data of subjects meeting specified exposure criteria.
 
Depending on pseudonymization versus anonymization of the data, subjects may need to sign an informed consent or have to be informed about the use of their pseudonymized data according to GDPR (General Data Protection Regulation). 
 
In both types of observational investigations, a clinical investigation plan is to be designed, and ethical approval is required. Good statistical power is needed to have a high-quality outcome of the study. 

Real-world data 

Real-World Data (RWD) are data relating to patient health status and/or the delivery of health care routinely collected from various sources such as: 

  • electronic medical records that capture patient encounters and treatments,  
  • claims and billing data reflecting health care utilization and costs,  
  • product and disease registries,  
  • personal devices such as smartwatches, social media, health apps, etc.  
The advantage of collecting RWD as part of PMCF is that it can cover the entire life cycle of the device in a study population that can represent the entire population as indicated in the Instructions for Use of the device and in all types of clinical sites. Analyses of real-world data can also provide insights into the device’s safety and performance.
 
However, depending on the region, privacy regulations such as GDPR can make the use of these data very difficult since anonymization and patient consent are crucial. Moreover, data quality impacts usability, bias affects generalizability, and combining data from different sources can be very complex, requiring expertise, the proper infrastructure and sufficient funding. 

Surveys, questionnaires, feedback from users 

Surveys are a valuable way of collecting PMCF data through a questionnaire sent to healthcare professionals, customers or patients. High-quality surveys can even bridge a substantial gap in clinical evidence. 

Surveys can either be patient-focused, where data collected are more specific around patients’ experience and outcomes. Here, the patient’s consent, ethics and GDPR will have to be taken into account. However, when designed to prospectively collect patient-reported outcomes and clinical outcomes, these surveys can be regarded as high-quality clinical data (rank 4) and may even be sufficient and a more cost-effective way, minimizing the need for clinical investigations. 

User surveys, focusing on the output provided by healthcare professionals (or lay users) of the medical device, usually provide fewer specific data and are generally ranked to be of lower quality data because of their retrospective character, not focusing on clinical outcomes but more on the experience with the device and the occasional device deficiency. 

Surveys also have their challenges for successful implementation, as the access to the end-user might be restricted, especially if a commercial distributor acts as an intermediary. The structure of the survey is of utmost importance to minimize bias; however, it cannot be too complex as this might result in a lack of responses.

Statistically designed surveys should ensure to provide sufficient evidence, but there should not be too many objectives as the loss of focus can result in skewed data; hence, experienced PMCF survey designers are needed to set up this type of activity. 

Requirements for PMCF activities 

PMCF activities must be meticulously planned and carefully designed to yield high-quality clinical data. Before planning, all clinical evidence should be reviewed, and gaps should be identified. Then, the following factors should be considered:  

  • Defining the objectives of the PMCF activities and aligning with the identified gaps, 
  • Sample size justification to create an appropriate plan (i.e., survey protocol, clinical investigation plan, etc.), 
  • Justification of the study design, 
  • Ethical requirements, if any, 
  • Data collection requirements, 
  • Statistical Analysis Plan, 
  • Reporting Requirements. 

PMS and PMCF documentation 

Post-market surveillance and PMCF studies should be thoroughly documented and trigger not only the update of the risk management documentation but also of the clinical evaluation and potentially the IFU and other applicable technical documents.  

PMCF plan 

In the PMCF plan, the manufacturer should document the methods in place to collect post-market clinical data on the device: 

  • The general methods and procedures 
  • A rationale of the appropriateness of these methods/procedures 
  • A reference to the relevant parts of the Clinical Evaluation Report 
  • The specific objectives addressed 
  • An evaluation of the clinical data related to similar or equivalent devices 
  • Reference to relevant CS/harmonized standards and relevant guidance 
  • Detailed and adequate time schedule of PMCF activities 

The PMCF plan is a required element of the post-market surveillance plan unless a justification is provided according to Annex III, Technical Documentation on PMS. Annex XIV Part B lists all required elements of the PMCF Plan 

PMCF report 

The PMCF report documents the findings of the PMCF activities. It should refer to each PMCF activity that was carried out, the planned design, the data collection methods and intended patient population and any deviations of these plans.

The report should then detail what the actual data sources were, how data have been collected and analyzed, considering safety and performance and clinical benefits of the medical device in scope.

Any new findings should be identified, and the impact on the other technical documentation should be assessed. Recommendations should be provided for any necessary actions (e.g., labeling updates, risk mitigation measures, etc.). 

This report is part of the clinical evaluation report and the technical documentation. Depending on the risk classification and type of the device, the PMCF report might be required to be updated annually. 

PMS Report or Periodic Safety Update Report (PSUR) – Which one should you prepare? 

Apart from the PMCF report, two types of documents, namely the PMS report and PSUR, can be prepared to report the general post-market surveillance data, depending on the device classification. The differences between the PMS report and the PSUR are outlined in MDR Articles 85 and 86.  

PMS report  

PMS reports are to be prepared for Class I devices and should present the post-market surveillance data and describe whether preventive or corrective actions were taken as a result of these activities. The PMS report is revised when necessary and provided to the competent authorities upon request, eliminating the requirement for submission.  

PSUR  

The PSUR is generally more comprehensive and focuses on safety and performance data over time. PSURs are specifically required for Class IIa to Class III devices and should provide: 

  • Post-market surveillance data 
  • Preventive and corrective actions taken 
  • Conclusions of the benefit-risk determination 
  • Results of the PMCF studies 
  • Device sales number 

For Class IIa devices, PSURs should be updated whenever necessary and at least every 2 years, while an annual update is required for higher risk devices (Class IIb and III). 

For Class III and implantable devices, PSURs must be made publicly available via EUDAMED, and the NB should also provide its evaluation on this electronic platform, offering transparency regarding the device’s safety and performance to healthcare providers and patients. For other device classes, the PSUR should only be provided to the NB and, upon request, to the competent authorities.   

How can we assist you during the post-market stage of your medical device? 

QbD Regulatory can help you to prepare the Technical Documentation, including risk management and post-market surveillance documents. 

QbD Clinical Medical Writing specializes in performing the required Clinical Evaluation for your device, including addressing the possible gaps in clinical evidence for your device in scope. This involves reviewing and analyzing all necessary literature and writing the clinical evaluation plan and report.

Further gaps in clinical evidence can be handled by setting up the right PMCF activities, by writing the proper PMCF plan, including the defined activities to collect more clinical data on the safety and performance of the device under evaluation to eventually produce a PMCF report and, hence an updated clinical evaluation report in due time. These documents are created following all applicable guidance and regulations. 

QbD Clinical Operations handles PMCF investigations from setup to execution and data analysis. Don’t hesitate to reach out to us for assistance, whether it is coaching, reviewing, or fully writing your technical documentation as required. 

 

Stay ahead in life sciences

Keeping up with the fast-paced life sciences industry doesn’t have to be overwhelming.

Our newsletter delivers the latest insights, industry updates, and expert content directly to your inbox, helping you stay informed and make smarter decisions.

Circles-banner-short

Discover more expert content

preview_image
Case study

Scalable MSAT validation support for seamless compliance & efficiency

Discover how QbD Group helped a global pharmaceutical company address fluctuating validation demands across multiple teams.
preview_image
Case study

Restoring manufacturing timelines through expert deviation management

Discover how QbD Group helped a leading biotechnology company overcome a backlog of deviations, restoring timely batch release through expert QA support.
preview_image
Case study

Guiding a global biotech company through ATMP facility qualification

Discover how QbD Group supported a global biotech company in commissioning and qualifying two ATMP therapy production sites, ensuring compliance and efficiency.
preview_image
Whitepaper

How to keep computerized systems in the operational phase

Ensure compliance and efficiency with best practices for maintaining computerized systems in the operational phase. Download our expert whitepaper now!
preview_image
Blog

What Makes Usability Testing Crucial for Near-Patient and Self-Testing Devices under IVDR?

It shouldn’t be a surprise that today, “Near-Patient Testing (NPT)” and...
preview_image
Blog

When does Annex XIV apply in Performance Studies, and what key documentation is needed for compliance?

In the European regulatory landscape, conducting performance studies for in...
preview_image
Blog

How to define your Clinical Performance Strategy?

1. Start with a clear intended purpose A strong clinical...